SECTION 2—COMPONENT REPLACEMENT
Replacing the Patient Outlet Filter
NOTE: For this procedure, refer to FIGURE 2.3.
NOTE: The patient outlet filter should be replaced according to the Transportable Oxygen
Concentrator Preventive Maintenance Record on page 23.
1. With the Transportable Oxygen Concentrator in the upright position, locate the
patient outlet filter.
2. Place fingers on patient outlet filter and turn counter clockwise to loosen (DETAIL
"A").
3. Remove patient outlet filter and o‐ring from unit (DETAIL "B").
4. Install new patient o‐ring onto unit.
5. Install patient outlet filter.
A. Hand tighten by turning clockwise until secure.
NOTE: Use only Invacare replacement Part Number 1164372 (Patient Outlet Filter Kit) on the
Transportable Oxygen Concentrator.
Shown
without
O-Ring
Replacement Parts
NOTE: Refer to the portable oxygen concentrator parts catalog at www.invacare.com for
component replacement parts.
™
SOLO
2
DETAIL "A"
Patient
Outlet
Filter
DETAIL "B"
FIGURE 2.3 Replacing the Patient Outlet Filter
Shown
with
O-Ring
12
Part No 1164898