SECTION 2—COMPONENT REPLACEMENT

Replacing the Patient Outlet Filter

NOTE: For this procedure, refer to FIGURE 2.3.
NOTE: The patient outlet filter should be replaced according to the Transportable Oxygen 
Concentrator Preventive Maintenance Record on page 23.
1. With the Transportable Oxygen Concentrator in the upright position, locate the 
patient outlet filter.
2. Place fingers on patient outlet filter and turn counter clockwise to loosen (DETAIL 
"A").
3. Remove patient outlet filter and o‐ring from unit (DETAIL "B").
4. Install new patient o‐ring onto unit.
5. Install patient outlet filter.
A. Hand tighten by turning clockwise until secure.
NOTE: Use only Invacare replacement Part Number 1164372 (Patient Outlet Filter Kit) on the 
Transportable Oxygen Concentrator.
Shown
without
O-Ring

Replacement Parts

NOTE: Refer to the portable oxygen concentrator parts catalog at www.invacare.com for 
component replacement parts.
SOLO
2
DETAIL "A"
Patient
Outlet
Filter
DETAIL "B"
FIGURE 2.3 Replacing the Patient Outlet Filter
Shown
with
O-Ring
12
Part No 1164898