7 Troubleshooting/Maintenance - User/Dealer
7.1
Troubleshooting Electrical
SYMPTOM
• Electric functions do not move.
Part No 1180159
7 TROUBLESHOOTING/MAINTENANCE - USER/DEALER
PROBABLE CAUSE
• End of stroke reached.
• Bed not plugged in.
• Pendant not functioning.
• Cables.
• Household fuse or breaker
blown.
USER SOLUTIONS
• Operate opposite button.
• Ensure power cord is plugged into
power source.
• Ensure pendant cable not
entangled/pinched.
• Ensure cables are not
entangled/pinched.
• Call the dealer for
repair/replacement of cables
and/or motors.
• Check household fuse/breaker
box.
59
DEALER SOLUTIONS
• Operate opposite button.
• Ensure power cord is plugged into
power source.
• Ensure motor and power supply
are plugged into bus cable.
• Ensure pendant cable not
entangled/pinched.
• Ensure pendant is correctly
connected to power supply. Refer
to Electrical Connections on
page 55
• Ensure motor cable(s) and power
supply are correctly connected.
Refer to Assembling the Drive
Shaft on page 51.
• Ensure cables are not
entangled/pinched.
• Call Invacare for instructions to
repair/replacement of cables
and/or motors.
N/A
®
Invacare
G5510 Bed