SECTION 7—TROUBLESHOOTING
GUIDE
SYMPTOM:
Alarm:
Short beeps, long
pause
Concentrator not
operating,
power switch On.
Beep......Beep......
Alarm:
Continuous
Concentrator not
operating, Power
Switch On.
Beeeeeeeep....
Part No 1141493
SECTION 7—TROUBLESHOOTING GUIDE
PROBABLE CAUSE:
Main Power Loss:
1. Power cord NOT
plugged in.
2. No power at outlet.
3. Tripped circuit
breaker.
System Failure:
1. Unit overheating due
to blocked air intake.
2. Insufficient power at
outlet.
3. Internal repairs
required
27
SOLUTION:
1. Insert plug into outlet.
2. Inspect house circuit
breakers or fuses. If
problem recurs, use a
different outlet.
3. Call Service Provider.
1a. Remove and clean
cabinet filters.
1b. Move oxygen
concentrator at least 7.5 cm
(3 in) away from walls,
draperies or furniture.
2. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
3. Call Service Provider.