7.0 Additional display screens in calibration mode
7.1
High resolution reading:
The normal resolution of the thermometer is shown to the tenth of a degree
(Celsius or Fahrenheit). If resolution to the hundredth of a degree is desired,
upon completion of a reading press and release the Timer button. The "CAL"
display will change to "hr" and the temperature reading will show the last 3 digits
of the reading to the hundredth of a degree. A reading such as 99.2F would, in
high resolution, be shown as 924 (99.24) or 917 (99.17) as applicable.
To return to the calibration check mode ready state and prepare
the thermometer to take a second reading, press and release the Measure
button.
Ambient temperature reading:
7.2
The ambient temperature as read by the thermometer can be
displayed next. Press and release the Timer button to display the ambient
temperature reading. The "hr" display will change to "At"
and display the ambient temperature to a tenth of a degree.
To return to the calibration check mode ready state and prepare
the thermometer to take a second reading, press and release the Measure
button.
A second ambient sensor (thermistor) is also used (the lower of the two ambient
temperatures is used by the thermometer) and can be accessed by pressing the
Timer button once more. The screen will display "th" alternating with the display.
Sensor temperature reading:
7.3
The temperature of the thermopile sensor of the thermometer can be displayed
next. Press and release the Timer button to change from the ambient
temperature reading to the temperature of the thermopile sensor. The display
will change "th" to "St" and display the temperature of the sensor to a tenth of a
degree.
NOTE: Dependent upon the actions taken prior to entering this display, the
temperature may remain steady at the heated level or may decrease until
equilibrium with ambient is reached.
To return to the calibration check mode ready state and prepare the
thermometer to take a second reading, press and release the Measure button.
Target temperature reading prior to Measure button press:
7.4
The target temperature reading immediately prior to pressing the Measure
button can be displayed next. Press and release the Timer button from the
sensor temperature reading to display the pre-button press target temperature
reading, this will change the "St" display to "Pb" and show the target
temperature reading prior to Measure button was pressed.
Note: this reading is taken with the sensor heater activated and does not
maintain or achieve the same accuracy as the reading following the Measure
button press. Do not evaluate the calibration of the thermometer using this
display.
To return to the calibration check m ode ready state and prepare the
thermometer to take a second reading, press and release the Measure button or
the Timer button.
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
C / F
Mem
7

Advertisement

loading