Place the probe firmly into the ear device port.
Wait approximately three seconds, press the thermometer Measure button, and watch
for the Measure light to flash
Leave the thermometer in the 9600 Plus Calibration Tester until you hear a beep.
Remove the PRO 6000 Thermometer from the 9600 Plus Calibration Tester and read the
temperature in the thermometer's display. If the temperatures are within +/- 0.2º C
(+/-0.4º F), when testing with a probe cover installed or +/- 0.1 C (+/-0.2 F) if a probe
cover is not installed, the thermometer is within calibration. Note: it is recommeded that
all calibration testing be performed with a probe cover installed.
If temperatures are outside +/-0.2º C, verify the test procedure was strictly observed and
repeat the calibration check 2 more times at this temperature. If any two subsequent test
are within the acceptable limits the test may be considered a pass. If all 3 readings are
out of calibration, see section 7.2.
To perform the calibration check in high resolution mode see section 8, part 8.1.
Wait one full minute before taking another reading with the same thermometer.
Repeated measurements in short sequence may cause higher readings. (Note: the
PRO 6000 must be left on a table and not kept in the hand during this time.)
Note: The probe cover must be replaced
with a new probe cover before each reading,
to remove any residual heat absorbed from
the 9600 Plus Calibration Tester.
Test all available thermometers for calibration verification at the current calibration
set point temperature. Proceed to the next calibration set point temperature; see
"Changing the calibration set point temperature" in section 5.0
The device will exit CAL mode after ten minutes of inactivity, or may be forced to
exit by holding both the C/F button and Mem button for 5 seconds. Upon exiting
calibration check mode the thermometer will return to sleep mode.