6.0 Interpreting results

6.1 The thermometer is within calibration range if the measured values are at:
Tenth degree
Accuracy range with probe cover
36.0°C ±0.2°C (96.8°F ±0.4°F)
38.5°C ±0.2°C (101.3°F ±0.4°F)
41°C ±0.2°C (105.8°F ±0.4°F)
Hundredth degree (high resolution)
The smallest significant digit is the tenths digit (0.X), therefore when interpreting data at the hundredths digit
(0.XX) the reading must be rounded to the nearest tenth prior to interpretation. As such, the accuracy ranges
are as follows:
Accuracy range with probe cover
36.0°C ±0.24°C (96.8°F ±0.44°F)
38.5°C ±0.24°C (101.3°F ±0.44°F)
41°C ±0.24°C (105.8°F ±0.44°F)
6.2 If results are outside the allowable tolerances, contact your local
Welch Allyn Customer Care Representative:
Go to www.welchallyn.com to find the contact information for technical support in your region.
Complaints or concerns:
Welch Allyn encourages all service centers to report customer complaints. This information is valuable for the
improvements of products.
All complaints where it is alleged that our product has caused or contributed to a death or a serious
deterioration in the state of health of a patient, user or other person, must be reported immediately to your local
Welch Allyn Organization.
A serious deterioration in the state of health can include a life-threatening illness or a condition necessitating
medical intervention to prevent permanent damage.
Also all cases should be reported where the circumstances indicate, that the incident was such that, if it
occurred again, it may lead to death or serious deterioration in health.