Notifications; Intellipap Travel Information; Intellipap Dc Operation; Battery Power - DeVilbiss IntelliPAP DV53 Standard Plus CPAP series User Manual

Positive airway pressure device
Hide thumbs Also See for IntelliPAP DV53 Standard Plus CPAP series:
Table of Contents

Advertisement

Available languages

Available languages

Clean Filter – The filter should be checked every 10 days for signs of dirt or wear so that it can be cleaned as needed.
This message helps to remind you to check it regularly.

notifications

Notifications identify conditions in your IntelliPAP that may require action by you or your equipment provider in order to
maintain a high level of therapy. If you see a message on the display that is not listed below, you may have the optional
DeVilbiss SmartLink Module attached to your device (Fig. F), which provides additional messages. Refer to the
documentation that was provided with the SmartLink Module.
Delay Running xx Minutes Left – While a comfort delay session is active, this message is flashed on the display every 5
seconds to let you know how much time is left in the delay.
Mask Leak – This message means the IntelliPAP has detected a large amount of airflow during therapy for at least 10% of
the time during your previous use session. This notification is displayed when the device is turned on. If this message is
displayed, put on the mask and adjust the headgear to ensure the mask is properly fitted to your face. Follow the mask
manufacturer's directions for adjusting the mask and headgear for proper fitting.
Mask Off – This message appears due to a poor mask fit or a removed mask. Check for leak around the mask seal and
make adjustments as necessary according to the mask manufacturer's instructions. This message will be displayed until
the high airflow problem is corrected. If the high airflow condition persists for about 20 seconds, the device will
automatically turn off if Auto-OFF is enabled.
Device Fault – If a device fault message is displayed, refer to Troubleshooting for instructions.

intelliPaP traVel inFOrMatiOn

International Power Changes
Your IntelliPAP is approved for sale and use in North America and is equipped with a universal power supply that is
automatically capable of accepting line voltages of 100 to 240V~ 50/60Hz. Simply contact your equipment provider for the
correct style power cord for the region in which you will be traveling (refer to Accessories/Replacement Items for the
correct part number) to adapt your unit to the country's power.
Altitude
Your IntelliPAP automatically compensates for changes in altitude between sea level and 9000 ft (2750 m). You do not
need to make adjustments to the device for changes in altitude.

intelliPaP Dc OPeratiOn

Battery Power

Your IntelliPAP can be operated from a 12V DC power source in a motor home, boat or other recreational vehicle using the
12V DC cable. Alternatively you can power your IntelliPAP using a 12V DC stand alone battery with a DC battery clamp on
adapter and 12V DC cable. All cables are available from your equipment provider (refer to Accessories/Replacement Items
for the correct part numbers). The stand alone battery connection uses clamps that attach to the terminal connections of
the battery.
NOTE–The heater for the IntelliPAP's humidification system will not operate directly from a 12V DC power source. If you
need humidification, non-heated humidification is still available. Alternately, instead of connecting12V DC power directly to
the IntelliPAP, you can pass the DC power through an inverter that will deliver AC power to the system. This will allow the
heater to function. The power rating of the inverter must be at least 200 watts @ 110V~ / 400 watts @ 220V~.
NOTE–The IntelliPAP will automatically switch power sources between AC and 12V DC if both sources are available. If
your unit is connected to both AC and 12V DC power sources, it will operate on AC power as long as AC power is present,
and automatically switch to 12V DC power if AC power is lost. When AC power is restored, it will return to the AC source.
You do not need to reset or adjust any settings when the power source switches, but remember that your heater will not
operate on 12V DC.

SuPPleMental OXYGen

WarninG-OXYGen uSe
Oxygen supports combustion. To avoid possible physical injury, do not smoke while using this device with
supplemental oxygen. Do not use this device near hot objects, volatile substances or sources of open flames.
Always turn on the IntelliPAP before turning on the oxygen source. Turn off the oxygen source before turning off the
IntelliPAP.
This device is not suitable for use in the presence of a flammable anesthetic mixture.
A-DV53
en - 9

Hide quick links:

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

Intellipap dv5 seriesIntellipap standart plus dv53Intellipap standart plus dv5hh

Table of Contents