Troubleshooting - DeVilbiss IntelliPAP DV53 Standard Plus CPAP series User Manual

Positive airway pressure device
Hide thumbs Also See for IntelliPAP DV53 Standard Plus CPAP series:
Table of Contents

Advertisement

Available languages

Available languages

4.
Hand wash the parts in a solution of warm water and mild dishwashing detergent.
CAUTION–Do not use any solutions that contain bleach, chlorine, alcohol, scented oils or other aromatic-based ingredients
to clean the humidifier components or tubing. These solutions may reduce the life of the product.
5.
Allow the parts to dry before reassembly.
CAUTION–Inspect the gasket seal before each use to ensure that it is properly seated in the groove around the entire
chamber.
NOTE– The disassembled chamber may also be washed once per week in a dishwasher on the delicate or glassware
cycle, top shelf only.
Humidifier Cradle - The humidifier cradle should be cleaned as needed.
1.
Turn off the heater and unplug from power source. Allow system to cool for at least 10 minutes.
2. D etach flow generator from the cradle; first remove water chamber then fully depress the flow generator release
button to separate (Fig 4).
NOTE–Flow generator release button must be pressed completely to separate.
3.
Use a soft, damp cloth to clean the surfaces of unit.
tubing
The air supply tubing should be cleaned daily. Remove the tubing from the device and mask. Use a mild detergent and
water to clean the inside of the tubing. Rinse the tubing and allow to air dry.
Mask and Headgear
Clean the mask and headgear according to the manufacturer's instructions.

trOuBleSHOOtinG

WarninG
Electric Shock Hazard - Do not remove the device cabinet; there are no user-serviceable internal components. The
cabinet should only be removed by a qualified DeVilbiss provider.
issue
Nothing shows on
the display.
The IntelliPAP does
not start when
breathing into the
mask.
en - 12
Possible cause
1. Device is not plugged in or the power cord
is not fully inserted.
2. Optional Humidifier-The flow generator is
not fully seated on humidifier cradle.
3. There is no outlet power.
1. The Auto-ON feature has been disabled.
2. There is no power to the device.
3. Breath is not deep enough for the Auto-ON
feature to detect.
4. You are using a full face mask that has an
anti-asphyxia valve.
5. The air supply port plug is missing or not
fully inserted.
6. The humidifier chamber is not fully engaged
into the cradle or is missing.
remedy
1a. Verify that power cord is firmly connected to
the IntelliPAP and the power outlet.
1b. If you are using a DC power source, make
sure the cable connections are secure. Verify that
battery is charged.
2. Refer to assembly instructions (Fig. 1-7).
3. Locate a suitable functioning power source.
1. Use the ON/OFF button to start and stop
device.
2. Verify that the power cord is firmly connected to
the device and the power outlet.
3. Breathe deeply in and out to start the device.
4. Auto-ON may not work because your exhaled
breath escapes through the open valve. Use the
ON/OFF button to start and stop the device.
5. Make sure the air supply port plug is fully
inserted into the unit.
6. Slide the humidifier chamber fully into the slot.
Verify the latching lever snaps into place.
A-DV53

Hide quick links:

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

Intellipap dv5 seriesIntellipap standart plus dv53Intellipap standart plus dv5hh

Table of Contents