DeVilbiss IntelliPAP DV53 Standard Plus CPAP series User Manual page 14

Positive airway pressure device
Hide thumbs Also See for IntelliPAP DV53 Standard Plus CPAP series:
Table of Contents

Advertisement

Available languages

Available languages

Humidifier
Humidity Output (in the operating flow range): . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ≥10 mgH
Size: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2.6" H x 6.3" W x 8.4" D (6.6 cm x 16.0 cm x 21.3 cm)
Weight: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.75 lbs. (0.794 Kg)
Power Consumption: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 Watts
DV5HH Electrical Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100-240VAC, 0.95 A
Heater Plate Thermal Cutout:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 305°F (152°C)
Power/Temperature Control: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Setting of 1 (minimum) to 10 (maximum)
Heater Plate Temperature: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84°F to 149°F (approx. 29°C to approx. 65°C)
Operating Temperature Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41˚F to 104˚F (5˚C to 40˚C)
Operating Humidity Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0% to 95% RH non-condensing
Operating Atmospheric Pressure Range:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70.0 – 106.0 kPA
Filter Specifications
Standard Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .> 3.0 micron particles
Optional Fine Particle Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .> 0.3 micron particles
Dynamic Short-term Pressure accuracy with and without DV5HH Humidifier (Per
iSO 17510-1:2007)
Waveform
Volume [mL]
Sinusoidal
cycle with I:E =
1:1
Maximum Flow-rate (per iSO 17510-1:2007)
Measured pressure at the patient
connection port (cmH
O)
2
Average flow at the patient connection port
(l/min)
Long Term Static Pressure Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +/- 0.5 cmH
additional Specifications
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Please contact homecare provider
Equipment classification with respect to protection from electric shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Class II
Degree of protection from electric shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Type BF Applied Part
Degree of protection against ingress of liquids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .IPXI Drip-proof vertical
Mode of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide
en - 14
Breath Rate [min-1]
500
10, 15, 20
3.0 cmH
Set Pressure
[cmH
3.0
7.0
12.0
16.0
20.0
Test Pressures
O
7.0 cmH
O
2
2
2.9
6.4
83.8
126.8
Pressure Accuracy
O]
[Pk-Pk cmH
2
12.0 cmH
O
16.0 cmH
2
11.0
15.0
137.4
130.0
O/l air
2
O]
2
0.5
0.5
1.0
1.0
1.0
O
20.0 cmH
O
2
2
19.0
113.8
O
2
A-DV53