Meter display
What it means
The InDuo
TM
meter has detected
that the temperature is below the
operating range. You cannot
perform a test until the InDuo
meter and the OneTouch
test strips reach a temperature
within the operating range of
43–111ºF (6–44ºC).
The InDuo
meter has detected
TM
that the temperature is above the
operating range. You cannot
perform a test until the InDuo
meter and the OneTouch
test strips reach a temperature
within the operating range of
43–111ºF (6–44ºC).
DISPLAY MESSAGES AND PROBLEM-SOLVING GUIDE
Action
Repeat the test after the
InDuo
OneTouch
have reached a temperature
TM
®
Ultra
within the operating range.
Repeat the test after the
InDuo
OneTouch
TM
have reached a temperature
®
Ultra
within the operating range.
77
TM
meter and
®
Ultra test strips
meter and
TM
®
Ultra test strips