The recommended environmental conditions:
indoors only
at an altitude up to 2000 m
at a temperature of 20...22 °C
at a relative humidity of 5 ... 80% (non-condensing)
6.

Instructions for checking the function of the system / quick start

When you open the package, please check to make sure all the necessary
parts are complete (see packing list) and nothing is damaged.
Check the electronics and the actuator for any visible damage:
The top and bottom plates of the actuator (if it does not have another shape)
should be parallel each to each other, and not contain any scratches.
If there is any damage to the system, please contact our local representative
immediately!
If the packaging material is damaged please confirm this with the shipping
company.
Before you switch on the system, please check:
that the main voltage supplied in your country is the same as the one
installed for the external power supply (not included in the shipment!).
that the secondary voltage (DC) matches with the amplifier.
Connect the power cable.
Connect the piezo element by using the LEMO connector.
Turn off closed loop (12V40SG / 12V40CLE).
Switch on the power supply. The red LED indicates that the amplifier is
working.
By turning the potentiometer to its maximum position, the actuator will reach
its maximum movement.
Offset is accomplished by the offset potentiometer or external offset (e.g. a
load). Because of the fast motion of the actuator a noise or crack can be
heard. This is normal and not a malfunction.
Turn the potentiometer into its minimum position.
12V40SG / 12V40CLE:
Turn on closed loop. Turn the potentiometer into the maximum position so
that the actuator does its maximum closed loop movement. Turn the
potentiometer into its minimum position and switch off closed loop.
Switch off the power supply. Ensure that the potentiometer is in its minimum
position.
The function check is now complete.