4-4-4.

DENSITOMETER BOARD

!
Caution
Any CIRCUIT BOARD that is replaced in the IMAGER must be returned to Kodak for rework or
disposed of properly. PRINTED CIRCUIT BOARDS contain lead.
1. Power down and remove the DENSITOMETER MODULE (4-4-1).
!
Warning
When the POWER CORD is plugged in, hazardous voltages are present in some areas of the
IMAGER. These voltages can cause severe injury or death.
2. Remove 4 HEX NUTS and WASHERS and remove the BOX from the DENSITOMETER (Figure 4-14).
3. Disconnect the RJ45 CONNECTOR from the DENSITOMETER.
4. Remove 5 ATTACHING SCREWS and remove the DENSITOMETER BOARD.
5. After reassembly and reinstallation of the DENSITOMETER MODULE, adjust the light source
reference level for the DENSITOMETER as described in paragraph 3-5 in Adjustments.
3
2005 July Rev. B 7F3318
Figure 4-14. Removing the DENSITOMETER
Section 4 - - Disassembly/Reassembly
4
2
8100- -73L
4-17