B. AC Voltage
Range
Resolution
4.000V
40.00V
400.0V
600V
Remarks:
Input impedance: 10M
l
Frequency response: 40Hz~400Hz.
l
Change to AC:
l
DCM-033:
Change to AC by using average response method. Input sine wave, then adjust
the reading until it is same as the effective value.
DCM-033T(Not Available)
Combine AC and True RMS response method. Input sine wave to adjust. Non
sine wave must follow the below data to adjust:
Peak factor: 1.4~2.0, add 1.0% on the stated accuracy
Peak factor: 2.0~2.5, add 2.5% on the stated accuracy
Peak factor: 2.5~3.0, add 4.0% on the stated accuracy.
Accuracy
1mV
10mV
100mV
1V
// less than 100pF
36
Overload protection
(1%+5)
(1.2%+5)
600V DC/AC