Preventive Maintenance Record - Invacare XPO100 Service Manual

Xpo series portable oxygen concentrator
Hide thumbs Also See for XPO100:
Table of Contents

Advertisement

SECTION 4—TROUBLESHOOTING AND MAINTENANCE

Preventive Maintenance Record

Model No. ________________________ Serial No._________________________
On Each Inspection
Record Date Of Service
Clean equipment surface
per Cleaning, Care and
Maintenance instructions*
Inspect XPO
cabinet and
2
accessories for damage**
Check for presence and
legibility of all labels
Annually, During Preventive Maintenance Schedule, Or Between Patients
Replace Patient Cannula
Clean or Replace Gross
Particle Filter*
Replace Inlet Filter*
Function Test (Refer to
Function Test on page 19)
Every 4,320 Hours Of Operation
Check Oxygen Purity*
As Required
Replace Internal Battery*
*NOTE: Refer to appropriate section of the Service or Owner's Manual for this procedure.
**NOTE: If damage is found, return unit to Invacare for repair.