Regulatory Listing - Invacare XPO100 Service Manual

Xpo series portable oxygen concentrator
Hide thumbs Also See for XPO100:
Table of Contents

Advertisement

TYPICAL PRODUCT PARAMETERS
Battery Duration (times are approximate):
Battery Recharge Time:
NOTE: Recharge time increases if battery is 
charging while unit is running.
Humidity
Temperature Range:
(All power sources)
Extended Temperature Range:
(Using AC or DC adapters)
Extended Temperature Range
(Using Internal Battery)

Regulatory Listing

ETL certified complying with:
XPO2™
Setting 1 = 3.5 hours
Setting 2 = 2.5 hours
Setting 3 = 2.0 hours
Setting 4 = 1.5 hours
Setting 5 = 1.0 hours
4 hours
Operating Humidity: 15% to 60% non condensing
Storage Humidity: up to 95% non condensing
Operating temperature: 41°F to 95°F (5°C to 35°C)
Storage temperature: -2°F to 140°F (-20°C to 60°C)
Operating Temperature:
95°F to 104°F (35°C to 40°C)
Continuous use - all settings
Operating Temperature:
95°F to 104°F (35°C to 40°C)
Continuous use - settings 1, 2 and 3
45 minutes (max) - setting 4
30 minutes (max) - setting 5
EN 55011:1998
CISPR 11: 2003
IEC 60601-1; 2nd ed. 2005
IEC 60601-1-2; 2.1 ed.
IEC 61000-3-2:2005
IEC 61000-3-3:2005
UL 60601-1, 1st ed.
CSA 601.1 M90
6
Part No 1150748