appendIX a - emc daTa
The ScanX is intended to be used within the electromagnetic environment specified below. The user of
the ScanX should assure that it is used in such an environment.
Immunity Test
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
Note: Ur is the a. c. mains voltage prior to application of the test level.
Page 28
Manufacturer's Guidance and Declaration -

Electromagnetic Immunity

IEC 60601
Test Level
± 6kV contact
± 6kV
± 8kV air
± 8kV
± 2 kV for power supply
lines
± 2kV
± 1 kV for Input/output
± 1kV
lines
± 1 kV line(s) to line(s)
± 1kV
± 2 kV line(s) to earth
N/A
<5% Ur (>95% dip in
<5% Ur - 10ms
Ur) for 0.5 cycle
40% Ur (60% dip in Ur)
40% Ur - 100ms
for 5 cycles
70% Ur (30% dip in
70% Ur - 500ms
Ur) for 25 cycles
<5% Ur (>95% dip in
<5% Ur - 5 s
Ur) for 5 s
3 A/m
3A/m
Compliance
Level
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic mate-
rial, the relative humidity should be at least 30%.
Mains power quality should be that of typical
commercial or hospital environment.
Mains power quality should be that of typical
commercial or hospital environment.
Mains power quality should be that of typical
commercial or hospital environment.
If the user of the ScanX requires continuous
operation during power mains interruptions in
the main power supply, it is recommended that
the ScanX be powered from an uninterruptible
power supply providing emergency power.
Power frequency magnetic fields should be at
levels characteristic of a typical location within
a typical commercial or hospital environment.
Electromagnetic
Environment - Guidance
Air Techniques, Inc.