Powering Down The System; Maintenance - Air Techniques ScanX Duo Instruction Manual

Table of Contents

Advertisement

poweRInG down sysTem
ScanX is designed to be left on continuously during the active day . At the end of the day, or whenever
desired, power down the system as follows:
1 . Switch the ScanX from ON to Standby by pressing and holding the membrane POWER switch
on the Keypad Control Panel for approximately two seconds, until the green LED above the
POWER switch extinguishes .
2 . Turn OFF the main power to ScanX by placing the main power switch on the Built-In Control/
Connector panel to the OFF (O) position .

maInTenance

Maintenance Procedures
ScanX is designed for many years of trouble-free operation . Maintenance as described herein is minimal .
IMPORTANT:
Cleaning ScanX
Turn off ScanX, disconnect the line cord from the Mains wall outlet and disconnect the computer
connection cable from ScanX before cleaning . Wipe the outside surfaces with a soft paper towel dampened
with a disinfectant solution or non-abrasive household cleaner . Be careful not to allow solvents TO RUN
OR DRIP into the unit . This could cause damage to ScanX . Allow to air dry before plugging in or turning
back on .
Disinfecting ScanX
No disinfection is required for ScanX .
Cleaning the Plate Guides
Wash the Plate Guides with a non-abrasive cleaner and water . Dry thoroughly .
IMPORTANT:
Note: Always use standard infection control procedures when handling devices that contact the patient.
Disinfecting the Plate Guides
There is no reason to routinely disinfect the Plate Guide unless contamination is suspected . If a Plate
Guide has touched a contaminated surface, it may be immersed briefly in a cold sterilant according to
sterilant manufacturers directions . After disinfection, clean and dry the Plate Guide using the instructions
above .
Cleaning the Plate Transport
Over time, small debris and dust can accumulate in the plate transport mechanism causing a loss in image
quality and possible damage to the PSPs .
To ensure optimal ScanX performance, the plate transport should be cleaned once per week using a new
ScanX Cleaning Sheet each time . Sheets can be purchased from your dealer . Refer to ACCESSORIES
page 24 .
Page 20
IMPORTANT:
Do not spray solvents or liquid directly on the scanner.
Plate Guides must not be put in an autoclave.
Never power down the system during a
scanning session.