KERN GAB Operating Instructions Manual page 14

Compact balance
Table of Contents

Advertisement

Procedure when adjusting:
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization.
Switch off the balance
Press and hold the TARE key and operate the ON/OFF key.
Release the TARE key. If display message deviates from (C1 7 / see on
right), this can be changed by pressing the MODE key. Afterwards
press the TARE key.
XX = model related display
41 = GAB / GAS 3K1DM
43 = GAB / GAS 15K5DM
41 = FAB 3K1DIPM
43 = FAB 15K5DIPM
Press the TARE key.
Press MODE key (value should be between 1000 and 25000).
Note: The display value is instable (extremely shock sensitive)
Press MODE key
(Ensure stable ambient conditions, do not load weighing plate)
resetting
Press
key, if value deviates from 5000
Place adjustment weight (See chpt. 1 technical specifications) carefully
in the centre of the weighing plate, press MODE key.
a value of c. 35000 appears, if deviation is in excess of 5000,
repeat sequence of operation.
Note: The display value is instable
(extremely shock sensitive)
Remove adjustment weight, adjustment complete.
Operate the adjustment switch at the bottom end of the balance (See
picture below)
adjustment will be stored. Turn off balance by
pressing the ON/OFF key.
14
Operation
/
42 = GAB / GAS 6K2DM
/
34 = GAB / GAS 30K10DM
/
42 = FAB 6K2DIPM
produces a value of c. 5000
Display
L003b
C1 7
C2 XX
C3
13842
5,010 kg
5,000 kg
CAL
35,000 kg
P-EP
GAB/GAS/FAB-BA-e-0720