11.3. List of commands
out commands: Setting parameters or temperature values.
Command
out_mode_05
out_mode_05
out_sp_00
in commands: Asking for parameters or temperature values to be displayed.
Command
version
status
in_pv_00
in_sp_00
in_mode_05
11.4. Status messages
Status messages
00 MANUAL STOP
01 MANUAL START
02 REMOTE STOP
03 REMOTE START
11.5. Error messages
Error messages
-01 LOW LEVEL ALARM
-05 WORKING SENSOR ALARM
-03 EXCESS TEMPERATURE WARNING
2
-07 I
C-BUS ERROR
-08 INVALID COMMAND
-09 COMMAND NOT ALLOWED IN
CURRENT OPERATING MODE
-10 VALUE TOO SMALL
-11 VALUE TOO LARGE
26
Parameter
0
1
xxx.xx
Parameter
none
none
none
none
none
Response of recirculating cooler
Stop the unit = R –OFF-.
Start the unit.
Set working temperature
Response of recirculating cooler
Number of software version (V X.xx)
Status message, error message (see page 26 )
Actual bath temperature.
Working temperature
Recirculating cooler in Stop/Start condition:
0 = Stop
1 = Start
Description
Recirculating cooler in „OFF" state.
Recirculating cooler in keypad control mode.
Recirculating cooler in „r OFF" state.
Recirculating cooler in remote control mode.
Description
Low liquid level alarm.
Working temperature sensor short-circuited or
interrupted.
High temperature warning. Starting at 75 °C
(no deactivation)
The return temperature soon reaches the switch-off
value of the high temperature warning function (85 °C)
Internal error when reading or writing the I
Invalid command.
Invalid command in current operating mode.
Entered value too small.
Entered value too large.
2
C bus.