Warning functions and safety installations
Excess temperature protection
Low liquid level protection
Alarm message
Excess temperature - Warning function
Overload protection
Classification according to DIN 12876-1
Environmental conditions according to IEC 61 010-1:
Use only indoor.
Altitude up to 2000 m - normal zero.
Ambient temperature: +5 ... +40 °C
Air humidity:
Max. rel. humidity 80 % for temperatures up to +31 °C,
linear decrease down to 50 % relative humidity at a temperature of +40 °C
Max. mains fluctuations of ±10 % are permissible.
The unit corresponds to Class I
Overvoltage category
Pollution degree
Caution:
The unit is not for use in explosive environment.
Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes) Class A
(industrial and commercial range).
10
85 °C - fixed value
float switch
optical + audible (permanent)
75 °C
for compressor and pump motor
class I
II
2