PRODIGY AutoCode® Blood Glucose Meter
Caution: The control solution ranges are
located on the back of the Prodigy® test strip
vial. They are not recommended target ranges
for your blood glucose.
Important Information:
• Use only Prodigy® control solutions with your
Prodigy® meter.
• Check the expiration date on the bottle.
Do not use if expired.
• Use within a period of 90 days from the date
that you first open it. Record the
discard date on the control solution bottle
when you first open it to serve as a reminder
to discard after 90 days.
• For in vitro diagnostic use.
Why Perform a Control Solution Test:
• To ensure that your meter and test strips are
working properly together.
• To allow you to practice testing without using
your own blood.
When to use:
• Once a week (to ensure that you continue to
have accurate results).
Owner's Manual
PRODIGY AutoCode® Blood Glucose Meter
• When you begin using a new vial of test strips
• When test strips have been exposed to extreme
environmental conditions.
• If you drop the meter.
• If you change the battery in the meter.
Storage and Handling:
• Store the control solution tightly closed at
temperatures below 86°F (30°C).
• Do not refrigerate. Do not store in your car.
Composition:
Low Solution 0.05
High Solution 0.19
30
Glucose Nonreactive Ingredients
(% w/v) (% w/v)
99.95
99.81
Owner's Manual