Reassembly - Fluke 867B User Manual

Graphical multimeter
Hide thumbs Also See for 867B:
Table of Contents

Advertisement

2. Remove the six screws attaching the case
halves. Detach the case halves.
3. Remove the two screws securing the shield as
shown in Figure 7-1. Then pry the shield loose
from two stanchions at the top of the circuit
board.
4. Gently pull up on the shield tab and rotate the
entire shield away from the circuit board.
5. Remove the three screws securing the circuit
board to the front half of the GMM.
6. Dislodge the circuit board by pushing in on the
Volts-Ohms input connector from the lower right
front of the GMM. (See Figure 7-1, Detail A).
7. Lift the entire circuit board up approximately one
inch, dislodging the slotted Battery Eliminator
connection.( See Figure 7-1, Detail B).
8. Rotate the circuit board 180 degrees.
9. Remove the fuse.
Note
For a more detailed drawing, see the
Service Manual (PN 689312).
User Maintenance

Reassembly

Generally, reassembly reverses the disassembly
procedure. Specifically, do the following:
1. Verify that the selector switch still points to OFF.
2. Place the case top face down on a non-marring
surface.
Note
At all times, avoid stressing the ribbon cable.
3. Grasp the board at the top and bottom so that it
can be smoothly lowered into position in the
case top.
4. As you begin lowering the board into position,
make sure the battery eliminator connection
(Detail B in Figure 7-1) slips into its slot snug
against the top case wall.
5. Gently press the board into position at the top
and bottom
6. Replace the three screws securing the board.
7. Rotate the shield back into place on the circuit
board.
8. Replace the case bottom. Tighten the six screws
(middle screws first) securing the case bottom.
7
Reassembly
7-3