Error Code
592
593
594
801
802
803
1000
1001
1002
1003
1004
1005
1006
1007
1008
65535
Error Code
701
702
703
704
706
707
708
709
710
711
712
713
82
Table 31. Ethernet/GPIB Configuration Errors (cont.)
Error Text
ESTB
ESRQ
ETAB
ENET Port number
ENET Server
ENET Timeout
Unknown function
Unknown range
Unknown CCO
Can't set NPLC
Can't set trigger
Can't get DCV Gains
Line Freq High
Line Freq Low
Write CCO failure
Unknown
Table 32. Calibration Errors
Error Text
CAL Secure Code Case Invalid characters in calibration secure code. The secure code
CAL Secure
CAL Secure code
CAL Secure code too
long
CAL Invalid step
number
CAL Reference
CAL Not calibrating
CAL Abort
CAL Sequence
CAL No measurement
CAL Overload
CAL Too far
Description
GPIB - Serial poll status byte lost.
GPIB - SRQ stuck in ON position.
GPIB - Table problem.
Port value is out of range (1024 to 65535).
Could not open the ethernet port.
Time-out occured while opening the ethernet port.
Unknown Function Selection.
Unknown Range Selection.
Unknown Calibration Constant.
Illegal/Unknown NPLC Selection.
Illegal/Unknown TRIGGER Selection.
Cannot get 1V/10V DC linearization constants.
AC Line frequency too high.
AC Line frequency too low.
Calibration Constant write failed.
Unknown Error.
Description
must contain only characters 0-9 and A-Z. No lower case or
other special characters are allowed.
Cal secured.
Invalid secure code.
Secure code too long.
Invalid calibration step number.
Cal reference value out of tolerance.
CAL? only works if you are calibrating.
Calibration aborted.
Calibration steps out-of-sequence.
No measurements taken during calibration.
Overload at input during calibration.
Calibration measurements out-of-tolerance.