Fluke 8845A Programmer's Manual page 64

Digital multimeter
Hide thumbs Also See for 8845A:
Table of Contents

Advertisement

8845A/8846A
Programmers Manual
TRIGger:SOURce(?) {BUS|IMMediate|EXTernal}
Description Sets the source from which the Meter will sense a measurement trigger.
Parameters <n>
Examples
Query
TRIGger:DELay <n>
Description Sets the delay between receiving a trigger and the beginning of
Parameters <n>
Examples
Query
TRIGger:DELay:AUTO <n>
Description Enables or disables automatic trigger delay. The amount of delay is set
Parameters <n>
Examples
Query
SAMPle:COUNt {<value>|MINimum|MAXimum}
Description Sets the number of measurements the Meter takes per trigger. This setting
Parameters <n>
Examples
56
= BUS
IMMediate
EXTernal
TRIG:SOUR BUS
TRIG:SOUR?
measurement cycle. If the command is used without a value for n, the
Meter automatically sets a delay time.
= 0 to 3600
MINimum
MAXimum
TRIG:DEL 14
TRIG:DEL?
TRIG:DEL? MIN
TRIG:DEL? MAX
based on selected function, integration time, and ac filter setting. When a
specific trigger delay is set using TRIG:DELAY <n> command, automatic
trigger delay is disabled. Setting is stored in volatile memory.
= ON or 1
OFF or 0
TRIG:DEL:AUTO ON
TRIG:DEL:AUTO?
is stored in volatile memory.
= 0 to 50,000
MINimum
MAXimum
SAMP:COUN 2450
Sets the Meter to expect a trigger through the
IEEE bus or upon execution of a *TRG
command.
Selects Meter's internal triggering system.
Sets Meter to sense triggers through the
trigger jack on the rear panel of the Meter.
Sets the Meter to expect a trigger through the
IEEE bus or upon execution of a *TRG
command.
Retrieves trigger source to which the Meter has
been set (BUS, IMM, or EXT).
Delay specified in seconds.
Delay set to 0 seconds.
Delay set to 3600 seconds.
Sets the trigger delay to 14 seconds.
Retrieves the amount of time set for a trigger
delay.
Returns 0 seconds
Returns 3600 seconds
Automatic trigger delay enabled.
Automatic trigger delay disabled.
Enables automatic trigger delay
Retrieves the state of the automatic trigger delay
( 0 = OFF, 1 = ON).
Number of measurements per trigger
Number of measurements per trigger set to 1.
Number of measurements per trigger set to
50,000.
Number of measurements per trigger set to
2,450.