Parameters <n>
Example
DATA:FEED?
Description Retrieves the state of the reading store setting.
Query
CALCulate:KMATh:MMFactor(?) <value>
Description Sets the value in the "m" register for the mx+b calculation.
Parameters <value> = 0 to ±999.999999
Examples
Query
CALCulate:KMATh:MBFactor(?) <value>
Description Sets the value in the "b" register for the mx+b calculation.
Parameters <value> = 0 to ±999.999999
Examples
Query
CALCulate:KMATh:MUNits(?) <name> = 3 characters using 'A' through 'Z'
Description Sets the units for the mx+b calculation.
Parameters <name> = A to ZZZ
Examples
Query
CALCulate:KMATh:STATe(?) <n>
Description Enables or disables the mx+b calculation.
Parameters <n>
Examples
Query
Programming the Trigger System
The Meter is designed to provide extensive control over the triggering of a single
measurement or series of measurements. One trigger can cause the Meter to take a single
measurement or multiple measurements up to 50,000 per trigger.
Operating the Meter remotely requires selecting one of three trigger methods: internal,
external, or bus triggering. Once a source is selected, the Meter must be in a "wait-for-
trigger" state before the trigger will start a measurement cycle.
Figure 4 shows the multimeter's triggering system.
54
= "CALCulate"
""
DATA:FEED RDG_STORE, "CALC".
DATA:FEED?
CALC:KMAT:MMF 37 Sets "m" register to 37.
CALC: KMAT:MMF?
CALC:KMAT:MBF 37 Sets "b" register to 37.
CALC: KMAT:MBF?
CALC:KMAT:MUN VOL
CALC: KMAT:MUN?
= ON or 1
OFF or 0
CALC:KMAT:STAT ON Enables mx+b calculations.
CALC: KMAT:STAT?
Enable the storing of measurements.
Disable the storage of measurements.
Returns CALC or "".
Retrieves the "m" register value.
Retrieves the "b" register value.
Up to three letters
Sets units to "VOL".
Retrieves the set units value.
Enables mx+b calculations.
Disable mx+b calculations
Retrieves the state of the mx+b calculation (0 =
OFF, 1 = ON).