Fluke 8845A Programmer's Manual page 63

Digital multimeter
Hide thumbs Also See for 8845A:
Table of Contents

Advertisement

Initiate Triggering
MEASure?
READ?
INITiate
Trigger Source
TRIGger:SOURce IMMediate
TRIGger:SOURce EXTernal
TRIGger:SOURce BUS
Front-panel "Single" key
Triggering Delay
TRIGger:DELay
Sample (*)
Annuniator
The Meter can be put into a wait-for-trigger state remotely by sending it a MEASure?,
READ?, or INITiate command. Once received by the Meter, it takes approximately 20
ms to setup the Meter and put it in to the wait-for-trigger state. Any triggers received
during this set-up time are ignored.
Using the Triggering Commands
INITiate
Description Sets the Meter to the wait-for-trigger state in which the next trigger from
the selected source triggers a measurement cycle. Up to 5,000
measurements are placed into the Meter's internal memory, where they can
be retrieved with the FETCh? command.
Examples
INIT
READ?
Description Sets the Meter in to the wait-for-trigger state where the next trigger from
the selected source triggers a measurement cycle. Measurements are sent
directly to the output buffer.
Examples
READ?
Idle
State
Wait-for-
Trigger
State
Delay
Measurement
Sample
Figure 4. 884XA Triggering System
Sets the Meter to the wait-for-trigger state.
Sets the Meter to the wait-for-trigger state.
Supported SCPI Commands
Sample
Trigger
Count # 1
Count # 1
Digital Multimeter
dae02.eps
55