Calibration Commands - Fluke 8845A Programmer's Manual

Digital multimeter
Hide thumbs Also See for 8845A:
Table of Contents

Advertisement

8845A/8846A
Programmers Manual
*PSC?
Description Returns the state of the clearing of the Status Byte and Standard Event
Examples
*SRE <enable value>
Description Enables or disables bits in the Meter's Status Byte enable register.
Parameters <n>
Examples
*SRE?
Description Retrieves the value in the Meter's Status Byte enable register.
Examples
*STB?
Description Retrieves the value in the Meter's Status Byte summary register. Although
Examples

Calibration Commands

CALibration? [{ON|OFF}]
Description After entering the proper code using the CAL:SEC:CODE command, this
Parameters <n>
Examples
62
Enable registers at power-up.
*PSC?
= 0 to 255
*SRE 48
*SRE?
like a serial poll, this command is processed as an instrument command.
Returned results are the same a when receiving a serial poll, but the SRQ
bit is not cleared if a serial poll has occurred.
*STB?
command causes the Meter to perform an adjustment using the values
supplied with the CAL:VAL command. See the 8845A/8846A Service
Manual for details on using this command.
= ON or 1
OFF or 0
CAL? ON
CAL?
Returns zero if clearing is disabled. Returns one
if clearing enabled.
Binary-weighted decimal value representing the
bits to set in the register. Although this register
has 8 bits and will accept a number from 0 to
255, only bits 3 through 5 are used.
Sets bits 4 and 5 to one and all other bits to zero.
Retrieves a binary-weighted decimal value
representing the bits set in the enable register.
Execute only one calibration step.
Execute a group (mode) of related calibration
steps at one time.
Execute one step of the calibration steps for the
calibration group specified by the CAL:VAL
command.
Returns +0 (good) or +1 (failed)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8846a

Table of Contents