Selecting Line Voltage

The 5725A arrives from the factory configured for the line voltage normally appropriate
for the country of purchase, or as specified at the time of purchase. Check the line power
label on the rear panel of the 5725A to verify that the line voltage matches local line
power. Figure 2-3 shows the location of the line power label and switches, and the
switch settings for each line voltage.
The 5725A can be set to operate from eight different nominal line voltages; each voltage
setting has a voltage tolerance of ±10% and a frequency range of 47 to 63 Hz. The line
voltage switches are located on the bottom left side of the rear panel.
To change the line voltage setting, set the line voltage selection switches to the correct
setting shown in Figure 2-3.
WARNING
GROUNDING
CONNECTOR IN POWER CORD
MUST BE CONNECTED TO
ENSURE PROTECTION FROM
ELECTRONIC SHOCK.
47-63 Hz
750VA MAX
!
FUSE
CHASSIS
GROUND
S2
CAUTION
FOR FIRE PROTECTION
REPLACE ONLY WITH A 250V FUSE
OF INDICATED RATING.
VOLTAGE
SELECTION
S2 S3 S4
100V
110V
115V
120V
S2 S3 S4
200V
220V
230V
240V
CAUTION
FOR FIRE PROTECTION
REPLACE ONLY WITH A 250V FUSE
OF INDICATED RATING.
VOLTAGE
FUSE
SELECTION
S2 S3 S4
100V
!
110V
F8A 250V
115V
(FAST)
120V
S2 S3 S4
200V
!
220V
F4A 250V
230V
(FAST)
240V
S3
S4
Figure 2-3. Line Power Label and Switch Location
FUSE
!
F8A 250V
(FAST)
!
F4A 250V
(FAST)
JOHN FLUKE MFG. CO., INC.
MADE IN U.S.A.
PATENTS PENDING
NO INTERNAL USER SERVICEABLE
PARTS REFER SERVICE TO
QUALIFIED SERVICE PERSONNEL
TO CLEAN FILTER
REMOVE FROM INSTRUMENT
AND FLUSH WITH WARM
SOAPY WATER
5700A CALIBRATOR
Installation

Selecting Line Voltage

2-7.
CURRENT
OUTPUT
HI
11A MAX
20V PK
LO
MAX
CALIBRATION
ENABLE NORMAL
aq4f.eps
2
2-7