Fluke 5725A Instruction Manual page 29

Amplifiers
Hide thumbs Also See for 5725A:
Table of Contents

Advertisement

DC Current
For fields strengths >1 V/m but <3 V/m, add 1% of
Resolution
Range
24 Hours
nA
220 µA
.1
2.2 mA
1
22 mA
10
µA
220 mA
.1
2.2A
1
[Note 1]
5725A Amplifier:
11A
10
330 + 470
Secondary Performance Specifications and Operating Characteristics
Included in Uncertainty Specifications
Stability ±1°C 24
Range
Hours
± (ppm output + nA)
V
220 µA
5 + 1
2.2 mA
5 + 5
22 mA
5 + 50
220 mA
8 + 300
9 + 7 µA
2.2A
± (ppm output + µA)
5725A:
11A
25 + 100
Notes:
Maximum output from 5700A terminals is 2.2A. Uncertainty specifications for 220 µA and 2.2 mA ranges are increased by 1.3 X when
supplied through 5725A terminals.
Specifications are otherwise identical for all output locations.
1. Add to uncertainty specifications:
±200 X I
2
ppm for >100 mA on 220 mA range
±10 X I
2
ppm for >1A on 2.2A range
2. Temperature coefficient is an adder to uncertainty specifications. It does not apply unless operating more than ±5°C from calibration
temperature.
3. Burden voltage adder is an adder to uncertainty specifications that does not apply unless burden voltage is greater than 0.5V.
4. For higher loads, multiply uncertainty specification by:
0.1 x actual load
1 +
maximum load for full accuracy
5. 5700A compliance limit is 2V for outputs from 1A to 2.2A. 5725A Amplifier may be used in range-lock mode down to 0A.
Minimum output: 0 for all ranges, including 5725A.
Settling time to full accuracy: 1 second for µA and mA ranges; 3 seconds for 2.2A range; 6 seconds for 11A range; + 1 second for range
or polarity change
Overshoot: <5%
Absolute Uncertainty
±5°C from calibration temperature
range
90 Days
180 Days
± (ppm output + nA)
45 + 10
50 + 10
55 + 10
45 + 10
50 + 10
55 + 10
45 + 100
50 + 100
55 + 100
± (ppm output + µA)
55 + 1
60 + 1
65 + 1
75 + 30
80 + 30
90 + 30
340 + 480
350 + 480
Temperature Coefficient
[Note 2]
0°-10°C
Compliance
10°-40°C
and
40°-50°C
± (ppm output + nA)/°C
1 + .40
3 + 1
1 + 2
3 + 10
1 + 20
3 + 100
3 + 1 µA
1 + 200
1 + 2.5 µΑ
A3 + 10 µA
± (ppm output + µA)/°C
20 + 75
30 + 120
Introduction and Specifications
1 Year
60 + 10
60 + 10
60 + 100
70 + 1
95 + 30
360 + 480
Burden
Maximum
Voltage
Load
Adder
For Full
Limits
[Note 3]
Accuracy
[Note 4]
±nA/V
10
.2
20k
10
.2
2k
10
10
200
10
100
20
2 µA
3
2
[Note 5]
4
0
4
Specifications
Relative Uncertainty ±1°C
24 Hours
90 Days
± (ppm output + nA)
24 + 2
26 + 2
24 + 5
26 + 5
24 + 50
26 + 50
± (ppm output + µA)
26 + .3
30 + .3
40 + 7
45 + 7
100 + 130
110 + 130
Noise
Bandwidth
Bandwidth
0.1-10 Hz
10-10 kHz
pk-pk
RMS
ppm output + nA
nA
6 + .9
10
6 + 5
10
6 + 50
50
9 + 300
500
12 + 1.5 µA
20 µA
ppm output + µA
µA
15 + 70
175
1
1-13

Advertisement

Table of Contents
loading

Table of Contents