Introduction

The Fluke 5725A Amplifier enhances the 5700A Calibrator in the ac voltage, ac current,
and dc current functions. The 5725A operates under complete control of the 5700A
through an interface cable supplied with the 5725A.
A diagram in the specifications tables at the end of this section illustrates the extended ac
volt-hertz product achieved by using a 5725A. Increased ac voltage load limits allow
using the 5700A Calibrator in systems with long cables.
Voltage output from the 5725A is available at the 5700A Calibrator front or rear binding
posts. This eliminates the need to move cables during a procedure that requires amplified
as well as standard calibrator outputs.
The 5725A front or rear panel OUTPUT binding posts are only for current output.
Extended-range ac and dc current is supplied through them. Since most meters with a
high current range use a separate high current input terminal, this configuration normally
eliminates the need to move cables during a procedure. If a single-point current output is
needed, the 5700A Calibrator can be configured to source all current outputs through the
5725A binding posts.
Enhancements to 5700A ac voltage output capability provided by the 5725A are as
follows:
Frequency limits at higher voltage increase to 100 kHz at 750V, 30 kHz at 1100V.
Load limits are to 70 mA for frequencies above 5 kHz, and to 50 mA for frequencies
less than 5 kHz.
Capacitive load limits are increased to 1000 pF.
Model 5725A operating functions and ranges are as follows:
AC voltage: 220 to 1100V rms up to 70 mA (50 mA < 5 kHz), 40 Hz to 30 kHz; 220
to 750V rms up to 70 mA, 30 kHz to 100 kHz
DC current: 0 to ±11A
AC current: 1 to 11A rms, 40 Hz to 10 kHz