Users can set system sounds and overtime of the backdrop lighting .
4.4.2.
4.4.2.2. 2. 2. 2. SpO2 Parameters
4.4.2.
4.4.2.
  Drop for Event (%): value, in %, to qualify as an SpO2 (Desaturation) Event.
  Minimum Event Duration (sec): value, in seconds, of SpO2 decrease that qualifies
as an SpO2 (desaturation) event.
  Desaturation Criteria Level (%): value, in %, below which the SpO2 level must
drop to be classified in a special category of statistics. This threshold is indicated on
the SpO2 graph by a red dotted line.
Pulse Rate Parameters
  Rate Change For Event (bpm): value, in beats per minute, to qualify as a pulse rate
event.
  Minimum Event Duration (sec): value, in seconds, of pulse rate change to qualify
as a pulse rate event.