Indicators
Alternating short
and long beeps
Description
Depleted Battery Alarm
This alarm occurs when
approximately two minutes
of battery life remains� The
battery symbol flashes
on-screen�
Alarm Silence Symbol
This symbol appears when
you press the Alarm Silence
button to silence the
audible alarm (beeps)�
Attention
This symbol appears
when corrective action is
required�
Battery Charging While
Almost Empty Symbol
This battery symbol
appears (with the charging
bars scrolling on the left
side) when the battery
is almost depleted and
recharging�
SimplyGo User Manual
What To Do
Replace or
recharge battery�
(Press Alarm
symbol to silence
the alarm�)
No further action
is needed�
Allow the device
to warm up� If the
alarm continues,
contact your
home care
provider�
No action needed�