Closure Commands For Ramp Segments - Memmert VO200 Operating Instructions Manual

Vacuum drying oven, pump module
Table of Contents

Advertisement

20.1 Closure commands for ramp segments

NEXT
SPWT (T)
SET-POINT WAIT
SPWT (P)
SET-POINT WAIT
SPWT (PT)
SET-POINT WAIT
LOOP
HOLD
END
The programme segments are linked together by the segment closure command. These commands
therefore control the programme sequence.
closure
command
ramp
°C
segment 1
spwt (t)
delayed
programme start
Follow-on with next programme segment.
Wait until the setpoint temperature is reached.
The oven only starts the next programme segment when the programmed
setpoint temperature has been reached, even if the programmed heating
time has already elapsed.
Wait until the set vacuum has been reached.
The oven starts the next programme segment only when the programmed
vacuum has been reached, even if the programmed segment time has
already elapsed or if the set temperature has already been reached.
Wait until temperature setpoint and vacuum setpoint have been
reached.
The oven starts the next programme segment only when the programmed
setpoint temperature and the programmed vacuum have been reached,
even if the programmed segment time has already elapsed.
Ramp repeat function.
The set programme is repeated after passing through all programmed
segments.
1-99 = repeats
cont = continuous repeat function
End of programme without switching off the heating; temperature and
all other settings (e.g. air valve) are maintained.
Programmende mit Abschaltung der Heizung und Belüftung des
Innenraums.
closure
command
ramp
closure
command
segment 3
ramp
spwt (t)
segment 2
next
segment 1
Ramp
closure
command
ramp
segment 4
next
segment 4
page 25
closure
command
ramp
segment 5
end
t=time
PERFECT

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VO200 and is the answer not in the manual?

This manual is also suitable for:

Vo500Pm200Vo400Pm400Pm500

Table of Contents

Save PDF