seca
Instructions for PC-based configuration
PC-based configuration is used for diagnosis or recalibration of all seca scales with modular electronics.
The following equipment is required:
•
PC with serial interface (RS -232)
•
seca software 'SERVA"
•
RS232 service set
consisting of:
•
RS-232-service module
•
Connecting cable PC/service module
•
Modular cable
•
Bus cable for DETF-LEM
•
Loading weights
•
For scales with DETF sensor at least 5 weight stages up to the MAX
•
For scales with DMS load cell at least 1 weight stage (=MAX)
Software Installation
If your 'SERVA'-software is in a ZIP-file, it has to be unpacked at first.
Execute the program's SETUP.EXE and follow the instructions displayed by the setup program.
Connecting and calibrating a scale
•
If the service module has not been connected to the PC yet, shut the PC down, connect the service
module to the serial interface and reboot the PC.
•
Switch off the scale!
•
For calibration, a connection between the service module and the measured-value
acquisition module of the scale to be calibrated must be established. For this purpose, a
modular cable must be connected to the scale's SeSAM bus.
17.06.03- Wint/Rei
Vorderseite / Frontside
Waage \ Scale
Anzeige / Display
1 (2)
page
10-02-01-004
68-60-00-002,
68-60-00-003
08-06-16-121
08-06-16-115
08-06-16-125
Rückseite / Backside
PC configuration program
Drucker
Printer
PC
30-34-00-672 b