Testing for battery drain
1. Insert the battery pill inside the hearing aid (example below).
2
2. Connect the battery pill into the red socket in the Affinity
test chamber.
3. If the test is included in the test sequence, it will be measured automatically.
Connect the battery pill
2
For more information on HIT440, please refer to the Affinity
Additional Information document.
Item No. 8105093 – 12/2013
Page 4 of 4