2. From Noah or from the Affinity suite AUD screen, enter the audiogram. The audiogram will be within the
range of normal/near normal.
3. Click on the REM tab.
4. Once in the REM screen, select the "Ear Level, FM only" protocol from the drop down list.
5. Place the FM transmitter microphone inside the test chamber, facing the reference microphone.
NOTE: The test box reference microphone is the active microphone
6. Close the lid of the test chamber.
7. Measure the MPO:
o
Click on MPO to access the MPO test.
o
Set the FM transmitter volume control to MAX.
o
Click Start to measure the Real-Ear Saturation Response.
o
If the MPO measurement is too high, adjust the FM output.
Item No. 8105097 – 12/2013