1.5.2 Power ON

With power applied, press On/ >O/T< /Off. All segments will appear briefly followed by Ohaus,
a software revision number, and then * 0.00 g.
The scale can operate on either AC Power, with the included AC adapter, or on
rechargeable battery. The scale automatically switches to battery operation if it is on and AC
power fails.
The battery must be fully charged for up to 12 hours before operation. Charge level is indicated
during charging. (See Table 1-3.) The scale can be operated during charging. The battery is
protected against over charging and the scale can remain connected to the AC power line.
Charge Status Indication Light
Allow 20 minutes warm-up time. The scale should be calibrated before use.
Note: If Auto-Off is set to ON, it must be set to OFF to allow 20 minutes warm-up time.
(See Menu Setup, next page.)
Note: To prevent water ingress during
use, insert rubber Seal into the input
jack recess.
Also note that prior to use the rear adjustable
feet should be opened at least a half turn.
There are tiny air vents on the inside of these
feet which allow the scale to "breathe," since it
is designed to be water-tight.

1.5.3 Power OFF

To turn the scale OFF, press and hold ON/>O/T</Off until OFF appears, then release.
Valor™ 2000 Series Service Manual
TABLE 1-3. BATTERY CHARGE STATUS
Off
Flashing Green
Flashing Yellow
Flashing Red
Rear Adjustable
Feet should be
opened at least
a half turn.
CHAPTER 1 GETTING STARTED
Charge Level
Fully Charged
70 – 100%
40 – 70%
0 – 40%
Figure 1-2. Waterproof seal on power
1-5
Ohaus Corporation www.ohaus.com
the internal
input jack.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

V21pw3V21pw6V21pw15