B.5 Service Span Calibration
When S.CAL appears, press YES to accept the Service Calibration menu,
or NO to advance to the End sub-menu selection.
When S.SPAN appears, press YES to accept the Service Span Calibration
menu or No to advance to the Service Linearity Calibration menu.
0 flashes. With no weight on the pan, press Yes to establish the zero
point. (Outlined characters represent a flashing display.)
- - C - - flashes while the zero point is established.
Service Span Calibration point flashes.
(The example shows the value for a Scale Capacity of 3 kg.)
Place the specified calibration weight on the pan and press Yes.
- - C - - flashes while the span point is established.
If span calibration was successful, the actual weight reading appears for
three seconds, followed by the Service Linearity Calibration menu item.
(The example shows a 3 kg span weight on the pan.)
Valor™ 2000 Series Service Manual
APPENDIX B SERVICE CALIBRATION
B-3
Ohaus Corporation www.ohaus.com
S.CAL
S.SPAN
--C--
--C--
3.000