Invacare MicroAir MA90Z User Manual page 24

Ma90z & ma95z series mattresses
Hide thumbs Also See for MicroAir MA90Z:
Table of Contents

Advertisement

4 OPERATION
iii.
Perform one of the following:
Cancel - Press the Turn Angle button to exit Fowler Transmitter Setup without linking to a transmitter.
Save - Wait 60 seconds without pressing a key.
MA95Z Power Unit - Perform the following steps:
i.
Press and hold the Hold and Max Inflate buttons.
ii.
AUTO FOWLER LEARN MODE will display.
iii.
PLACE BED IN FLAT POSITION will display.
iv.
CONTINUE -> SELECT will display.
v.
Press the Select button.
vi.
PLACE BED IN FOWLER/UPRIGHT POSITION will display.
vii. TRANSMITTER DETECTED will display.
viii. The control unit will beep.
ix.
STAND BY will display.
Mode Button (MA90Z Power Unit)
1.
Press the Mode button (
Max Inflate Mode - In this mode, the mattress inflates rapidly to maximum firmness (pressurized to 35 ± 6 mmHg). A series of beeps sound
every three minutes as a reminder that the Max Inflate mode is active. After 30 minutes, the Max Inflate mode deactivates and the control
unit defaults to the previous setting.
It takes 5-10 minutes for the MA90Z mattress to inflate fully (inflation time depends on size of mattress).
Max Inflate mode can be manually disengaged by pressing the Max Inflate button. This will deactivate the Max Inflate LED.
It is recommended that Max Inflate setting be used during patient ingress/egress, patient wound care, patient turning or
patient cleaning.
Low Air Loss Mode - In this mode, the mattress goes into on-demand low air loss relief mode.
®
Invacare
MA90Z Series Mattresses
) to select the Max Inflate Mode or the Low Air Loss mode.
24
Part No 1148139