Diode Test; Continuity Check - IDEAL 61-497 Instruction Manual

Digital multimeter
Hide thumbs Also See for 61-497:
Table of Contents

Advertisement

Available languages

Available languages

(4) Resistance
61-497
Range
Resolution
400.00Ω
0.01Ω
4.0000KΩ
0.1Ω
40.000KΩ
400.00KΩ
4.0000MΩ
100Ω
40.00MΩ
10KΩ
61-498
Range
Resolution
1000.00Ω
0.01Ω
10.0000KΩ
0.1Ω
100.000KΩ
1000.00KΩ
10.0000MΩ
100Ω
40.00MΩ
10KΩ
< 100dgt rolling.
[1]
Input Protection:
1000V
Maximum Open Circuit Voltage: Approximate 2.5V
Maximum Test Current:
Approximate 0.1mA
Note: For best measurements, use the REL

(5) Continuity Check

61-497
Range
Resolution
400.0Ω
0.1Ω
61-498
Range
Resolution
1000.0Ω
0.1Ω
Input Protection:
1000V
Maximum Open Circuit Voltage: Approximate 2.5V
Maximum Test Current:
Approximate 0.1mA
Continuity Threshold:
Approximate 50Ω
Continuity Indicator:
2KHz Tone Buzzer
Accuracy
± (0.2% + 30d)
10Ω
± (0.3% + 30d)
± (1.0% + 30d)
[1]
± (1.5% + 30d)
[1]
Accuracy
± (0.05% + 30d)
± (0.025% + 30d)
10Ω
± (0.3% + 30d)
± (1.0% + 30d)
[1]
± (1.5% + 30d)
[1]
or 1000V
rms
DC
AC
function to compensate for offsets.
Accuracy
± (0.2% + 3d)
Accuracy
± (0.05% + 3d)
or 1000V
rms
DC
AC
23

(6) Diode Test

61-497
Range
Resolution
2.000V
61-498
Range
Resolution
0.400V ~ 0.800V
Input Protection:
1000V
Maximum Open Circuit Voltage: Approximate ± 2.5V
Maximum Test Current:
Approximate 0.5mA
(7) Capacitance
61-497
Range
Resolution
4.000nF
1pF
± (1.2% + 40d)
40.00nF
10pF
400.0nF
100pF
4.000µF
1nF
40.00µF
10nF
400.0µF
100nF
4.000mF
1µF
40.00mF
10µF
± (1.2% + 40d)
61-498
Range
Resolution
4.000nF
1pF
± (1.2% + 40d)
40.00nF
10pF
400.0nF
100pF
4.000µF
1nF
40.00µF
10nF
400.0µF
100nF
4.000mF
1µF
40.00mF
10µF
± (1.2% + 40d)
< 50dgt rolling.
[1]
Input Protection:
1000V
Note: For best measurements, use the REL
Accuracy
1mV
± (1.5% + 2d)
Accuracy
1mV
± (1.5% + 2d)
or 1000V
rms
DC
AC
Accuracy
Measuring TIme
[1]
± (1.2% + 20d)
0.7 sec.
± (0.9% + 2d)
± (1.2% + 20d)
3.75 sec.
7.5 sec.
[1]
Accuracy
Measuring TIme
[1]
± (1.2% + 20d)
0.7 sec.
± (0.8% + 2d)
± (1.2% + 20d)
3.75 sec.
7.5 sec.
[1]
or 1000V
rms
DC
AC
function to compensate for offsets.
24

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

61-498

Table of Contents