For future sevicing work, please document any alterations to the factory default
settings.
Enter any alterations in the appropriate box.
General notes:
– Switch positions that have not been allocated a function are indicated in the
middle part of the display by a bar ( - ).
– The programming level is called up via the f position.
It is shown as E.. in the display.
– When processing medical devices, any changes made to programmes or
dispensing systems must be documented in a log book kept with the machine
(in accordance with the Medical Devices Directive). The machine's cleaning
and disinfecting standards must be re-validated.
The following applies to all programmable functions:
To activate programming mode
^ Turn the programme selector to f.
^ Switch off the machine.
^ Press and hold in ß and 6 at the
same time, and, whilst doing so,
press the I-0 switch.
To save values and exit from programming mode
^ Press 6.
^ Press 6 again.
To exit from programming mode without saving
^ Switch the machine off with the I-0
switch.
Programmable functions
The current programme version P.. will
appear in the display. The p / 6
indicator light comes on.
SP will appear in the display.
The alteration is now saved.
The alteration will not be saved.
35