SR-Series Manual
2. Press the
3. Once the LCD shows "8888" tap the
times until a random code displays.
4. Press
by the required calibration weight (see chart).
5. Add the required weight to the hook. The display will
show pass and return to normal weighing mode.
6. Remove the weight and press and hold the
turn the scale off. Calibration is complete!
IX. Troubleshooting
Problem: Scale will not turn on
Solution: Check the batteries
Problem: Display will not stabilize
Solution: Make sure the weight is stable on the
hook and not shaking or swinging
Problem: Displayed weight is inaccurate
Solution: Perform a calibration adjustment
X. Error Codes
1. OUTZ - Zero setting has shifted. Perform calibration
adjustment. If the problem persists, the load cell may
be permanently damaged due to overloading.
2. EEEEE - Maximum capacity exceeded. Remove the
extra weight to avoid damaging the load cell.
3. LLLLL - Load cell has been overloaded.
key to turn the scale on.
again. The display will flash "CAL" followed
key five
key to