Dc Mv/Ac Mv Measurement - Aktakom ??M-1130 User Manual

Digital multimeter
Table of Contents

Advertisement

АМM-1130

DC mV/AC mV Measurement

The testing chart is shown in the Figure 3-3. The voltage measurement range is of 10μV~220mV and the measurement
methods are as follows:
1.
Turn on the power switch and set the rotary switch to the position
Insert the red testing line into the VΩHz end and the black testing line into the COM end.
2.
3.
Press the FUNC. key to select DCmV measurement modes.
4.
When performing DCmV or AcmV measurement, connect the red probe to the positive polarity of the measured
voltage and the black probe to its negative polarity.
5.
Read the measured value from the display screen. If OL displaying on the meter, it indicates the measured voltage
exceeding the range of the meter and it is necessary to remove both the red and black probes from the measured circuit
immediately.
6.
In this mode, it is null to press the RANGE key
Notes:
a)In case of probe hanging in the air, the voltage inducted by the testing line may cause unstable readings on the display screen,
but that will not affect the accuracy of measurement.
b)The meter is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter may display
a residual reading between 1 and 30 counts. It will not attect stated accuracy, but if using REL to offset this reading may produce a
much larger constant error in later measurements.
mV
.
19

Advertisement

Table of Contents
loading

This manual is also suitable for:

Amm-1130

Table of Contents