Radiant Innovation TH809 Operating Instructions

Radiant Innovation TH809 Operating Instructions

Ear-type thermometer

Advertisement

Quick Links

TH809 Ear-Type Thermometer
Thank you for purchasing the thermometer. This thermometer is
designed with an advanced infrared and ambient temperature
compensation technology for instantaneous self-diagnosis and
accurate temperature measurements.
Operating Instructions:
1. Gently squeeze the opposite ends of the thermometer to
pull off the probe cap. Do not use force to remove the cap.
2. Always use a new and undamaged probe cover.
Make sure the ear canal is clean.
3. Install Probe Cover
1) Place a new probe cover on the connection ring. (See Figure 1)
NOTE:Make sure to place the "Adhesive Side" of probe cover "Upward."
2) Align the probe with the center of probe cover. Insert the probe into the probe cover on the connection ring.
(See Figure 2)
3) Push the connection ring until the "Click" sound. This means the probe cover has been installed successfully.
NOTE:The probe cover detector is optional. If the probe cover did not connect firmly, the " " icon
will flash on the LCD screen. Please check the setting of the probe cover again.
Figure 1
Proper installation of the probe cover ensures accurate measurements.
Warning:Keep the probe covers and connection ring away from children.
4. Press "ON/MEM" button to power on. The thermometer is ready for use after the ear icon stop flashing and two short
beep sound.
5. Gently pull the ear back to straighten the ear canal and snugly position the
probe into the ear canal, aiming towards the membrane of the eardrum to obtain
an accurate reading. ( Fig.4-1)
6. Measuring the ear temperature:Use the index finger to trigger. Press the "Scan"
button until you hear a long beep. ( Fig.4-2 ).
7. Power off:This device will automatically shut down after 1 minute pending to extend battery life.
NOTE:
a. It is recommended that you measure 3 times with the same ear. If the 3 measurements are different, select the
highest temperature.
b. To avoid the risk of cross contamination, please clean the probe according to "Cleaning and Storage" section after
each use.
c. Clinical repeatability: 0.20°C (<1 year old), 0.14°C (1~5 years old), 0.14°C (>5 years old)
Figure 2
Figure 4-1
Figure 4-2
Switching between Fahrenheit
(°F)
and Celsius
In "Power Off" mode, press and hold the "SCAN" button, then press the "ON/MEM" button for 3 seconds, icon "°C"
will be switched to icon "°F". You can also use the same process to change the LCD display from °F to °C.
Memory Function:
If the reading of the thermometer is within normal temperature range between 34
when the thermometer is off, the last measurement data is saved into memory. Press the "ON/MEM" button
again to see the temperature stored.
Cleaning and Storage:
The probe is the most delicate part of the thermometer.
Use with care when cleaning the lens to avoid damage.
Storage temperature Range: It should be stored at room temperature between -20~+50°C, RH≦85%
Keep the unit dry and away from any liquids and direct sunlight.
The Probe should not submerge into any liquids.
** If device is accidentally used without probe cover, clean the probe as follows:
a. Please use the cotton swab with Alcohol (70% concentration) to clean the lens( on the inside of the probe).
b. Allow the probe to fully dry for at least 1 minute.
NOTE:Please check the device if damaged once it falls. If you can't make sure of it, please send the
complete device to the nearest retailer for recalibration.
Holding the thermometer too long may cause a higher ambient temperature reading of the
probe. This could make the body temperature measurement lower than usual.
Changing the Battery:
This device is supplied with one lithium cell (CR2032 x 1).
Open the battery cover: Insert a pointed object into the battery cover pick hole. At the same time, use thumb to remove battery
1
cover.
Flip the battery out with a small screw driver
(See Figure 1)
2
on the left side and press the right side of the battery down until you hear a "click".
The positive (+) side Up and the negative (-) side pointed Down.
Figure 1
Specifications:
Temperature measurement range: 34~42.2°C (93.2~108°F)
Operating temperature range: 10~40°C (50~104°F)
Storage temperature Range: It should be stored at room temperature between -20~+50°C , RH≦85%
Transportation temperature shall be less than 70°C, RH≦95%
Compliance with ASTM E1965-98, EN12470-5:2003 Clinical thermometers-Part 5:Performance of infra-red ear
thermometers(with maximum device),IEC/EN60601-1-2(EMC),IEC/EN60601-1(Safety) standards.
Accuracy:+/-0.2
o
C (0.4
o
F) during 35.5~42°C (95.9~107.6°F), +/-0.3
This mode converts the ear temperature to display its "oral equivalent." (according to the result of the clinical evaluation)
(°C)
:
o
o
C to 42.2
C (93.2
** Replace the probe cover after each use to
ensure an accurate reading and avoid cross
contamination.
Insert the new battery under the metal hook
(See Figure 2)
3
Replace the battery cover
(See Figure 3)
4
Figure 2
Figure 3
o
C (0.5
o
F) for other range.
o
o
F to 108.0
F),

Advertisement

loading

Summary of Contents for Radiant Innovation TH809

  • Page 1 TH809 Ear-Type Thermometer Switching between Fahrenheit (°F) and Celsius (°C) Thank you for purchasing the thermometer. This thermometer is designed with an advanced infrared and ambient temperature In “Power Off” mode, press and hold the “SCAN” button, then press the “ON/MEM” button for 3 seconds, icon “°C”...
  • Page 2 Please read the instructions for use BF type applied part Medical Technology Promedt Consulting GmbH Add: Altenhofstrasse 80, D-66386 St. Ingbert, Germany Radiant Innovation Inc. Ref No.:022010 Tel: +886 (3) 5644185 Fax: +886 (3) 5644170 E-mail:service@radiantek.com.tw Http://www.radiantek.com.tw Add: 1F, No.3, Industrial East 9...
  • Page 3 Symbol Descriptions Guidance and manufacturer’s declaration – electromagnetic immunity The CE mark and Indicates this The TH8xyz series is intended for use in the electromagnetic environment specified below. The Notified Body device is subject to customer or the user of the TH8xyz series should assure that it is used in such an environment. Registration the Waste Numbers, the...
  • Page 4 Guidance and manufacturer’s declaration – electromagnetic immunity Recommended separation distances between portable and mobile RF communications equipment and the The TH8xyz series is intended for use in the electromagnetic environment specified below. The ME EQUIPMENT or ME SYSTEM customer or the user of the TH8xyz series should assure that it is used in such an environment. IEC 60601 test Compliance Electromagnetic environment –...