THP0FU Non-contact Forehead Thermometer
Intended use:
The infrared forehead thermometer can measure the infrared heat generated by
forehead and surrounding tissues, to reflect patient's body temperature accurately.
Intended operator:
At least 11 years old (5 years intensive reading experience), no maximum.
 Temperature measurement range: 34~42.2°C (93.2~108°F)
Specifications
 Operating temperature range: 10~40°C (50~104°F), 15%~85% RH
 Storage temperature range: It should be stored at room temperature between
-20~+50°C , RH≦85%
Transportation temperature shall be less than 70°C, RH≦95%
 Atmospheric pressure: 800~1013 hPa
 Comply with ASTM E1965-98, EN ISO 80601-2-56, IEC/EN60601-1-2(EMC),
IEC/EN60601-1(Safety) standards, ISO10993, RoHS.
 Accuracy:
±0.2°C (0.4°F) within 35~42°C (95~107.6°F), ±0.3°C (0.5°F) for other range.
 Fever indication, memory and °C/°F switch function
 Battery: one lithium cell battery (CR2032 *1pcs).
 Battery life: around 3,000 continuous readings.
 Expected Service Life: 4 years
 This thermometer converts the forehead temperature to display its "oral equivalent."
(according to the result of the clinical evaluation to get the offset value)
 Enclosure Rating: IP22
 Dimensions: 136.3 x 48.7 x 40.4 mm
Weight: 65.3 grams including battery
The device should not submerge into any liquids and expose it to direct moisture.
There is no gender and age limitation for using the infrared thermometer.
This is not an AP or APG product.
Functions
The thermometer has been designed for practical use. It's not meant to replace a visit to the doctor.
Please also remember to compare the measurement result to your regular body temperature. Please
Forehead temperature
consult with doctor if you have health concerns.
 Please see the "Use of the thermometer" section to learn how to measure the body temperature.
If the thermometer detects a temperature ≧37.5°C (or 99.5°F) under forehead mode, three short
Fever indication
beep sound will follow one long beep sound to warn the user for potential fever.
There are total 25 sets of measurement condition for body temperature.
Memory locations
 When power on, press the ON/MEM button to see the temperature records with
In "Power Off" mode, press and hold the "SCAN" button, then press the ON/MEM button for 3
°C / °F switch
seconds, icon "°C" will be switched to icon "°F". You can also use the same process to change the
LCD display from °F to °C.
Use of the thermometer
Note: The device must stay in stable ambient (room) temperature for 15 minutes before operating.
1. Always make sure the probe lens is clean without any damage and the forehead is clean.
Warning: Choking from swallowing small parts and batteries by children or pets is possible, please keep small parts and batteries at places where
children and pets can't reach.
2. Power on:
Press the ON/MEM button (see figure 1).
3. Measuring body temperature on the forehead:
Press the ON/MEM button to power on the device. Forehead mode is the default mode. You can see the
and hear two beep sounds (see figure 1). In this mode, you can hold the thermometer within 1 cm from the central forehead (Fig. 2) and press
the "Scan" button to get the forehead measurement. The time consuming for measurement might be 1 second. After each forehead
measurement, wait
icon stop flashing to be ready for next measurement.
Points for attention:
a. Forehead temperature is displayed in oral mode. This mode converts the forehead temperature to display its "oral-equivalent" value.
b. Before the measurement, please stay in a stable environment for 5 minutes and avoid exercise, bath for 30mins.
c. Remember to keep the forehead area clean and away from sweat, cosmetics and scar while taking temperature.
d. The "Clinical Bias" is -1.4 ~ -1.7°C.
e. The "Limits of Agreement" is 0.98.
f. The "Repeatability" is 0.20°C.
4. Power off:
4.1 Device will automatically shut off if left idle for more than 1 minute to extend battery life.
4.2 Manually power off the device by pressing the ON/MEM button.
5. Communication Interface: 3-stage phone jack, the temperature reading will be outputted with "standard UART communication format" after finishing measurement.
central
icon.
icon on the screen
Figure 1
Figure 2

Advertisement

loading

Summary of Contents for Radiant Innovation THP0FU

  • Page 1 THP0FU Non-contact Forehead Thermometer central Intended use: The infrared forehead thermometer can measure the infrared heat generated by forehead and surrounding tissues, to reflect patient’s body temperature accurately. Intended operator: At least 11 years old (5 years intensive reading experience), no maximum.
  • Page 2 Warning: No modification of this equipment is allowed. Please read the instructions for use BF type applied part Radiant Innovation Inc. Medical Technology Promedt Consulting GmbH Http://www.radiantek.com.tw Add: Altenhofstrasse 80, D-66386 St. Ingbert, Add: 1F, No.3, Industrial East 9...
  • Page 3 Guidance and manufacturer’s declaration – electromagnetic emissions The THP0FU is intended for use in the electromagnetic environment specified below. The customer or the user of the THP0FU should assure that it is used in such an environment. Electromagnetic environment – guidance...
  • Page 4 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. Guidance and manufacturer’s declaration – electromagnetic immunity The THP0FU is intended for use in the electromagnetic environment specified below. The customer or the user of the THP0FU should assure that it is used in such an environment.