Guidance and manufacturer's declaration – electromagnetic immunity
The TH8xyz series is intended for use in the electromagnetic environment specified below. The
customer or the user of the TH8xyz series should assure that it is used in such an environment.
IEC 60601 test
Immunity test
level
Electrostatic
discharge
6 kV contact
(ESD)
8 kV air
IEC 61000-4-2
Electrical fast
2 kV for power
transient/burst
supply lines
1 kV for
IEC 61000-4-4
input/output lines
1 kV line(s) to
Surge
line(s)
IEC 61000-4-5
2 kV line(s) to earth
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
interruptions and
40 % UT
voltage
(60 % dip in UT)
variations
on
for 5 cycles
power
supply
70 % UT
input lines
(30 % dip in UT)
for 25 cycles
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 5 sec
Power frequency
(50/60 Hz)
magnetic field
3 A/m
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
Compliance
Electromagnetic environment –
level
guidance
Floors should be wood, concrete
6 kV
or
ceramic
tile.
If
contact
covered with synthetic material,
8 kV air
the relative humidity should be at
least 30 %.
Mains power quality should be that of
Not applicable
a typical commercial or hospital
environment.
Mains power quality should be that of
Not applicable
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
TH8xyz series
requires
Not applicable
operation
during
interruptions, it is recommended that
the TH8xyz series be powered from
an uninterruptible power supply or a
battery.
Power frequency magnetic fields
should be at levels characteristic of a
3 A/m
typical
location
in
commercial or hospital environment.
The TH8xyz series is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the TH8xyz series can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the TH8xyz series as recommended below,
floors are
according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
continued
power
mains
a
typical
Recommended separation distances between
portable and mobile RF communications equipment and the
ME EQUIPMENT or ME SYSTEM
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
d = 1,2
d = 1,2
0,12
0,38
1,2
3,8
12
m
800 MHz to 2,5 GHz
d = 2,3
0,12
0,23
0,38
0,73
1,2
2,3
3,8
7,3
12
23