TH90K(E) Multi-function Infrared Thermometer
Specifications
Temperature measurement range:
Ear/Forehead mode: 34~42.2°C (93.2~108°F), Surface mode: -22~80°C (-7.6~176°F)
Operating temperature range: 10~40°C (50~104°F)
Storage temperature Range: It should be stored at room temperature between -20~+50°C,
RH
8 5%
Transportation temperature shall be less than 70°C, RH
Comply with ASTM E1965-98, EN12470-5:2003 Clinical thermometers-Part
5:Performance of infra-red ear thermometers(with maximum device),
IEC/EN60601-1-2(EMC), IEC/EN60601-1(Safety) standards.
Accuracy:
For ear mode: +/-0.2°C (0.4°F) within 35.5~42°C (95.9~107.6°F),
+/-0.3°C (0.5°F) for other range.
For forehead mode: +/-0.2°C (0.4°F) within 36~39°C (96.8~102°F),
For surface mode: +/-0.3°C (0.5°F) within 22~42.2°C (71.6~108°F),
Fever alarm & Memory under " Ear/Forehead mode "
°C / °F Switch mode
Battery: one lithium cell battery (CR2032 *1pcs)
This thermometer converts the ear/forehead temperature to display its "oral equivalent."
:LED Backlight (optional)
LED Backlight will automatically turned on after measurement, and automatically turned
off after 5 seconds
Do not use this device for Live & Dead decision or Safety related applications. Please
consult with doctor if you have health concerns.
There is no gender and age limitation for using infrared thermometer.
This is not an AP or APG product.
Functions
Suitable ambient temperature is important for the baby and patient, the thermometer always help you recognize the
room temperature.
When in power off, room temperature will always be shown on the screen with
Room Temperature
Please see the "Use of the thermometer" section to learn how to get the correct room temperature.
How to turn off Room Temperature: When in power on, press and hold the "ON/MEM" button for 5 sec. until
"OFF" will be shown on the screen to release the button.
The thermometer has been designed for practical use. It's not meant to replace a visit to the doctor. Please also
Ear Temperature
remember to compare the measurement result to your regular body temperature.
Please see the "Use of the thermometer" section to learn how to measure the body temperature.
The thermometer has been designed for practical use. It's not meant to replace a visit to the doctor. Please also
Forehead Temperature
remember to compare the measurement result to your regular body temperature.
Please see the "Use of the thermometer" section to learn how to measure the body temperature.
The surface mode shows the actual, unadjusted surface temperatures, which is different from the body temperature.
Surface Mode
It can help you to monitor if the object temperature is suitable for the baby or patient, for example the baby's milk.
Please see the "Use of the thermometer" section to learn how to measure the object temperature.
If thermometer detects a body temperature
Fever Alarm
sound to warn the user for potential fever.
When you get a new temperature reading in ear mode, the last reading will be shown on the screen (at top right
Last Reading
corner) with the
There are total 25 sets memories for ear measurements.
Memory Locations
When in power on, Press the "ON/MEM" button to see the temperature stored with
In "power off" mode (The LCD only shows the room temperature). Press and hold the "SCAN" button, then
°C / °F Switch
immediately press and release the "ON/MEM" button. After 3 seconds, icon "°C" will be switched to icon "°F". (You
may use the same way to switch back to "°C")
Use of the thermometer
1. Always make sure the probe is clean, and without damage.
2. Power on: Press "ON/MEM" button.
3. Mode selection:
3.1 Ear mode: Press the "ON/MEM" button, push up the raised brim on the
sliding panel on the above of LCD screen(see figure 4) until
hearing a click sound ,the mode of thermometer is ear mode.
In this mode, please press the "SCAN" button (see figure 5)
to get the body temperature by ear measurement.(see figure 1)
3.2 Forehead mode: Press the "ON/MEM" button , push back the raised brim
on the sliding panel on the above of LCD screen(see figure 6),
the mode of thermometer is forehead mode.
In this mode, you can get the body temperature by forehead
measurement.(see figure 2)
+/-0.3°C (0.5°F) for other range.
others +/-4% or +/-2°C (4°F) whichever is greater.
icon.
9 5%
3 7.5°C, there will be a long beep sound followed by three short beep
icon.
icon.
Ear Mode
Figure 1
Figure 4
Figure 5
Figure 6
Figure 2
Figure 3
Forehead Mode

Advertisement

loading

Summary of Contents for Radiant Innovation TH90K

  • Page 1 TH90K(E) Multi-function Infrared Thermometer Specifications Temperature measurement range: Ear/Forehead mode: 34~42.2°C (93.2~108°F), Surface mode: -22~80°C (-7.6~176°F) Operating temperature range: 10~40°C (50~104°F) Storage temperature Range: It should be stored at room temperature between -20~+50°C, ≦ 8 5% Transportation temperature shall be less than 70°C, RH ≦...
  • Page 2: Important Notes

    3.3 Surface mode: After power on, press and hold the “ON/MEM” button, and press the “SCAN” button one time for “Surface” mode to see icon on your LCD display. In this mode, you can get the target surface temperature. (see figure 3) 4.
  • Page 3: Troubleshooting

    If interference occurs, please increase the separation further or turn off the electrical equipment in the vicinity. For detailed description regarding the EMC precautions you need to take, please contact our authorized representative or us on address as given below. Radiant Innovation Inc. Http://www.radiantek.com.tw Medical Technology Promedt Consulting GmbH Add: 1F, No.3, Industrial East 9 Road, Science-Based Industrial Park, Add: Altenhofstrasse 80, D-66386 St.
  • Page 4 Symbol Descriptions Indicates this device is subject to the Waste The CE mark and Notified Body Electrical and Electronic Authorized Registration Numbers, the Equipment Directive in the European Union. representative in the requirement of Annex II from Medical To protect the environment, dispose of European community Device Directive 93/42/EEC are met.
  • Page 5 Electrostatic Floors should be wood, concrete or ceramic tile. If floors are 6 kV contact 6 kV contact discharge (ESD) covered with synthetic material, the relative humidity should be at 8 kV air 8 kV air IEC 61000-4-2 least 30 %. Electrical fast transient/burst 2 kV for power supply lines...