I. SCALE OPERATION
Place the scale on a horizontal flat surface.
1. Press [ ON/OFF ] key.
2. Wait until "0.0" or "0.00" is displayed.
3. Place the object(s) on the weighing platform.
4. By pressing the [ MODE ] key, you can switch
between weighing units.
5. The removable covers can also be used as
weighing trays.
II. TARE WEIGHING
1. Turn on the scale as described above.
2. Place a tray or container on the scale platform.
3. Press the [ TARE ] button. "0.0" will be displayed.
4. You can now add items to the container or tray to
obtain just the weight of the contents.
5. When the container is removed, a negative number
will be displayed. Press [ TARE ] again to zero the
scale.
III. CALIBRATION
This scale has come pre-calibrated at the factory. Only
re-calibrate if you are experiencing trouble weighing.
SC-501 uses a 100g and 2x 200g calibration weight
SC-2kg uses a 1kg and a 2kg calibration weight
1. Turn the scale OFF.
2. For model SC-2kg: While holding the [ MODE ]
key, tap the [ ON/OFF ] key to turn the scale on then
release both keys.
For model SC-501: While holding the [ TARE ] key,
tap the [ ON/OFF ] key to turn the scale on then
release both keys.