Whip Mix Pro Press 100 Operation Manual page 37

Pro series
Table of Contents

Advertisement

Pro Series Furnace Operations Manual
2. Muffle Burn In
DO NOT FORGET TO PLACE THE FIRING TRAY
ON THE LIFT TABLE!
A. Press the Change Programs button.
B. Turn the Select Knob right to select ADD A
PROGRAM. Press the Next button.
C. Select program 0. Press the Next button.
D. The furnace will ask for a porcelain type. Turn
the Select Knob to select SPEC. Press the
Next button.
E. The furnace will tell you to enter a program name.
Turn the Select Knob and press the Next
button after every letter. For blank spaces merely
press the enter button without turning the Select
Knob. Enter the name BURN_IN.
0 SPEC – BURN IN
ENTRY TEMP 500C
Figure 56
F. Enter an entry time of 1 minute by turning
the Select Knob. Press the Next button.
G. Turn the Select Knob to select an entry
temperature of 500C. Press the Next button.
H. The furnace will ask for a soak temperature,
press the Next button to skip this item.
I.
The furnace will ask for a soak time,
press the Next button to skip this item.
J. Turn the Select Knob to select a heat rate of
40C/Minute. Press the Next button.
0 SPEC – BURN IN
HEAT RATE
Figure 57
K. Turn the Select Knob to enter a final temperature
of 1000C. Press the Next button.
0 SPEC – BURN IN
FINAL TEMP
Figure 58
L. Turn the Select Knob to select a hold time of
15 minutes. Press the Next button.
0 SPEC – BURN IN
HOLD TIME
Figure 59
40C/M
15:00
37
M. The furnace will prompt you to enter an SP CL
TIME, press the Next button to skip.
N. Turn the Select Knob to select a temperature of
1000C to lower the table. Press the Next button.
0 SPEC – BURN IN
LOWER TABLE 1000C
Figure 60
O. The furnace will ask you to enter a cool time, press
the Next button to skip. This will automatically
enter a cool time of 0:00.
0 SPEC – BURN IN
COOL TIME
Figure 61
P. Turn the Select Knob to select NO vacuum.
Press the Next button.
0 SPEC – BURN IN
USE vACUUM
Figure 62
Press the Run Program button.
Q.
Turn the Select Knob to select Program
R.
0 SPEC – BURN IN.
SELECT PROGRAM?
0 SPEC – BURN IN
Figure 63
Press the Special Functions button.
S.
Turn the Select Knob to CALIBRATE OVEN.
T.
Press the Next button.
Turn the Select Knob to the password number
U.
260. Press the Next button.
The program will begin to run immediately and will cycle
until the Stop Program button is pressed. Leave the
furnace to cycle this program for eight hours. When
you return press the Stop Program button.
Fire a chip of porcelain to test the firing temperature of the
new muffle. Adjust the calibration as described in Chapter
Four — Running a Program. Adjust furnace calibration
to match your porcelain. Run Vacuum Test, see page 38.