ReSound SY312 User Manual page 33

Receiver-in-ear hearing instrument
Hide thumbs Also See for SY312:
Table of Contents

Advertisement

technical data for micro/mini Rie—HP receiver
Models: AL960, AL760, AL560, ALT960 [-DR]
AL961, AL761, AL561, ALT961 [-DRW]
Reference test gain (60 dB SPL input)
HFA
Full-on gain (50 dB SPL Input)
Max
HFA
Maximum output (90 dB SPL input)
Max
HFA
Total harmonic distortion
500 Hz
800 Hz
1600 Hz
Equivalent input noise (w/o noise reduction)
Frequency range (DIN 45605)
Attack and release (ANSI RTG -7 dB)
Attack
Release
Current drain
Battery size 10A
Battery size 312
64
Input/Output Response
2cc Coupler
Open
Closed
Frequency: 2000 Hz
35
35
dB
57
58
dB
49
49
dB
117
117
dB SPL
112
112
dB SPL
0.8
0.7
%
1.1
1.0
%
0.9
0.8
%
26
26
dB SPL
100–7150 100–7140 Hz
12
12
ms
70
70
ms
Data in accordance with
ANSI S3.22–2003; Supply
1.2
1.2
mA
Voltage 1.3 V, 2cc coupler
1.4
1.4
mA
Maximum Output (OSPL 90)
2cc Coupler
Input (dB SPL)
Full-On and Reference Test Gain
2cc Coupler
Frequency (Hz)
Open configuration
Closed configuration
Full-on gain
50 dB SPL input
Reference test gain
60 dB SPL input
Frequency (Hz)
65